430
Power/Electronics Replacement
Replacing the main board
7.
If an MIO card is installed, remove the card and the support bracket as
described below:
a.
b.
c.
Figure 430-3
4 of 26
Loosen the two spring-loaded captivated screws on the back of the
instrument and slide the card out.
Remove the MIO support bracket.
Remove the three screws on the back of the instrument and the one
screw in the standoff on the side of the instrument.
Removing the card and three screws on the back of the GC
Mainframe
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001