Agilent Technologies 6890 Series Service Manual page 240

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

The baseline drifts (upward) significantly during an oven program
• If the oven temperature is increasing dramatically during a run (for exam-
ple, from 50 to 350°C) a change of between 10 and 15 pA is normal. Howev-
er, if you suspect that the baseline drift is excessive, heat the inlets and
oven to a temperature above 300°C for 60 minutes to eliminate excess
baseline drift during oven programs.
• Verify that the detector insulation is not cracked or damaged.
High detector baseline of GC at room temperature
• Moisture in the detector can cause the baseline to be at 10s or even 100s
of pA when the detector is at a low (such as room) temperature. Set the
detector temperature to 150°C with the detector gases on. The baseline
should drop below 1 pA in approximately 10 minutes.
The signal baseline does not fall below 3 pA when the bead voltage is 0
• The ceramic insulators may be dirty. The insulators must be very clean
for NPD performance to be satisfactory.
Jun 2001
Nitrogen Phosphorus Detector (NPD)
Correcting NPD hardware problems
Detectors
Agilent 6890 Gas Chromatograph Service Manual
320
9 of 32

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents