Replacing The Detector Interface Board - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

320
Nitrogen Phosphorus Detector (NPD)
Replacement procedures
14. Restore gas pressures and leak test the fittings.

Replacing the detector interface board

The NPD detector has an interface board on the main board.
WARNING
Before proceeding, turn off the detector and any detector gases at their
supply. Then turn off the main power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
1.
Remove the right side cover and the top electronics cover.
2.
Unlock and unclip the electrometer ribbon cable from the NPD interface
board.
3.
Disconnect the bead drive cable from the board.
28 of 32
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001