250
Volatiles Interface
Replacement procedures
14. Remove the screw at the top of the module as shown using a T-20 Torx
screw driver.
Figure 250-13
15. Unlock the folded ribbon cable and remove from the connector.
Figure 250-14
16 of 30
Removing the EPC inlet mounting screw.
Folded ribbon
cable
The ribbon cable
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Remove screw
Jun 2001