Automatic sampler, 6890A GC
This RS-232C serial port (labeled sampler on the back of the GC) is
preconfigured to connect the GC to an Automatic Liquid Sampler.
Cable pinouts, 6890 GC to Automatic Liquid Sampler
Part no. G1530-60600
5
9
6
1
Connector 1 — 9 pin (female)
1— DCD
6 — DSR
2 — Rx
3 — Tx
4 — DTR
5 — Ground
7 — RTS
8 — CTS
9 — RI
Jun 2001
G1530-60600
1
Cabling Configuration and Electronics
Agilent 6890 Gas Chromatograph Service Manual
Cable Electronics
Automatic sampler, 6890A GC
2
1
6
9
5
Connector 2 — 9 pin (female)
4 — DTR
3 — Tx
2 — Rx
6 — DSR
1 — DCD
5 — Ground
8 — CTS
7 — RTS
1020