Maintaining An Ecd Detector; Ecd Bakeout - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Maintaining an ECD detector

ECD bakeout

If your ECD baseline is noisy or the display frequency is too high (i.e., ≥100),
you should perform a thermal cleaning (also called a "bakeout") of the
detector. Before performing a bakeout, verify that the carrier supply gas and
flow system are leak- and contaminant-free.
Caution
Detector disassembly and/or cleaning procedures other than thermal should
be performed only by personnel trained and licensed appropriately to handle
radioactive materials. Trace amounts of radioactive
during these other procedures, causing possible hazardous exposure to β- and
x-radiation (bremsstrahlung).
WARNING
To prevent possible hazardous contamination of the area with radioactive
material, the detector exhaust vent must always be connected to a fume hood,
or otherwise vented in compliance with the latest revision of Title 10, CFR,
Part 20, or with state regulations with which the Nuclear Regulatory
Commission has entered into an agreement (USA only). For other countries,
consult with the appropriate agency for equivalent requirements.
1.
Record the ECD "Output" value from the GC display. If the number is
equal to or greater than 100, continue with this procedure.
2.
Turn off the anode purge and makeup gas flows.
3.
Remove the column from the detector.
4.
Cap the bottom of the makeup gas adapter with a blank column ferrule
and column nut.
5.
Set the makeup gas flow rate between 50 and 60 mL/min. Set the detector
temperature between 350 and 375°C.
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Electron Capture Detector (ECD)
Maintaining an ECD detector
63
Ni may be removed
340
23 of 24

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents