1225 Main Board, 6890N
General board layout
The test points on the G1530-60200 main board are shown in Figure 1225-1.
Some additional parts are included to assist in locating the test points.
+10 V test point
Figure 1225-1
Jun 2001
Ground
-10V REF
GND
Battery
Main board test points and battery, 6890N
Electrical
Agilent 6890 Gas Chromatograph Service Manual
F4
F5
+24V
-24V
+15V
-15V
+5V
GND
Main test point group
+24V ±10%
–24V ±10%
+15V ±5%
–15V ±5%
+5V
±2%
GND
1 of 6