Removing the detector cover

Removing the detector cover
The top cover protects the detectors, valve box, and valve assembly.
If your detector cover mounts in a bracket on the GC oven:
1.
Tilt the front cover up.
2.
Squeeze the clip on the right hinge, pull the clip toward you and pivot it up.
Figure 410-4
3.
Push the metal pin behind the clip to the left.
4.
Slide the detector cover to the right and lift it off.
5.
To replace the top cover:
a.
b.
c.
If your detector cover rests on a post on the electronic cover
4 of 20
Removing the detector cover
Slide the plastic post on the left side of the cover into the hole on the
injection port fan cover.
Line up the hole in the tab on the right side of the cover with the
metal pin and push the pin to the right, through the hole.
Pivot the clip back down and snap it back over the metal pin.