Agilent Technologies 6890 Series Service Manual page 713

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

1225
Main Board, 6890N
General board layout
Check the +10 V test point to determine if detector, A/D, and D/A circuitry
is operating.
If the instrument appears dead and there is no 24 V voltage:
1.
Check AC power
2.
Check fuses at F4 and F5.
If thermal shutdown occurs or if the detectors are not working:
1.
Check the –15 V voltage. If not present, A/D functionality will not work.
2.
Unplug the detectors and the EPC board.
3.
Recheck the –15 V voltage. If it returns, the problem may be in the
detector or EPC board.
2 of 6
Electrical
Agilent 6890 Gas Chromatograph Service Manual
Jun 2001

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents