Agilent Technologies 6890 Series Service Manual page 222

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

Figure 310-12
Caution
Make sure you are properly grounded with an ESD strap before continuing.
5.
Unlock the detector manifold's ribbon cable from the EPC board and
detach the connector. The adjacent ribbon cable may have to be removed
as well.
Caution
Always hold the manifold by its support bracket to avoid damaging board
components.
6.
Slide the manifold from its slot.
Jun 2001
Type 2 manifold
Type 1 manifold
Removing the detector flow manifold
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
Replacement procedures
Screw
310
15 of 24