Agilent Technologies 6890 Series Service Manual page 86

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

7.
Lay the EPC module on the lab bench and use an electronic leak detector
to locate the leaky component on the module.
Figure 220-12
8.
If the leaky component is serviceable, such as vent/inlet fittings (see
Figure 220-13 ) swap it out. Otherwise, replace the EPC module.
Jun 2001
To gas supply
Leak testing the EPC module on the lab bench
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed Inlet
Leak testing—EPC and manual inlets
Ribbon cable from
leak test kit
220
17 of 20

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents