Performing the leak test—EPC inlets
Note
Be sure to complete all of the preparation steps on pages 20 and 21 before
continuing.
1.
Set the inlet to "Split Mode."
2.
Configure the column as 0 length.
Press
[Config]
column of the "Dim" field.
Jun 2001
Leak testing—EPC and manual inlets
Press
[Mode/Type]
[
] or [
Column 1
Config
Inlets
Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet
] [
] and enter "0" in the first
Column 2
210
21 of 30