Removing the electrometer

WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off any detector gases at their supply. Then turn off the main
power switch and unplug the power cord.
1.
Remove the right side cover of the instrument.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
2.
Disconnect the electrometer's ribbon cable from the detector's interface
board.
3.
Remove the collector block (three Torx T-20 screws).
4.
Remove the two Torx T-20 screws and the corresponding clamps on the
electrical interconnect.
5.
Remove the Torx T-20 screw from each end of the electrometer and
remove the electrometer. (Do not remove the screw on the top of the
electrometer that holds the cover on.)
Collector block
Clamps
Interconnect
Figure 310-11
Jun 2001
Electrometer
Removing the FID electrometer
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID)
Replacement procedures
310
13 of 24