Correcting Leaks - Agilent Technologies 1100 Series Reference Manual

Thermostatted column compartment
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Column Compartment

Correcting Leaks

When required
Tools required
1 Remove the front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the column switching valve for leaks
4 Replace the front cover.
Column switching valve
Column
Leak sensor assembly
Waste outlet
Figure 15
62
If a leakage has occurred at the heat exchanger or at the capillary connections
or at the column switching valve
Tissue
Wrench 1/4 – 5/16 inch for capillary connections
and correct, if required.
Possible Leak Areas
1100 Series TCC Reference Manual