Replacement Procedures

Sample Line Replacement

Sample Line Replacement

Tools required

The following tools are required to perform the replacement procedure described following
section. Allow approximately 5 to 10 minutes to complete this process.
3/16 inch open-end wrench
Torx T-10 screwdriver

Replacement procedure for Micro GC sample line

1 Shut down the GC, and allow time for it to cool to ambient room temperature.
2 Disconnect the GC power plug from the mains outlet.
3 Disconnect any sample gas feed lines attached to the inlet port.
4 Remove the front cover as described in
5 Use a 3/16 inch wrench to unscrew the sample line nut from the front of the channels.
Agilent 990 Micro GC User Manual
Removing the front cover
on page 94.