Attach Quickswap To The Msd Transfer Line - Agilent Technologies QuickSwap G3185B Installation And Setup

Table of Contents

Advertisement

Overview
Restrictors are not labeled individually because of their size and because they will be
C A U T I O N
exposed to high vacuum. Instead, the containers are labeled and fitted with
color-coded caps (Red = 92 µm, Blue = 100 µm, Black = 110 µm, Green = 120 µm). To
avoid mixing them up, only remove one restrictor at a time and replace the cap
immediately.

3. Attach QuickSwap to the MSD Transfer Line

After selecting a restrictor, install it in the MSD transfer line and attach
QuickSwap.
1 The QuickSwap plate has three fittings, as shown in
To purge gas supply
Figure 7
20
If using one of the checkout methods provided, review it to verify the
flow rate is compatible with your MSD. If not, see the instructions on
modifying methods in the
plumbed for purge gas. The other two are used to connect to the MSD
transfer line and the column.
Sealed,
no connection
Fitting 1
Connection identification
Reference
Manual.
Column
connection
Fitting 2
Fitting 3
MSD transfer
line connection
QuickSwap Installation and Startup
7. Fitting 1 is
Figure