Verify Line Voltage, Voltage Settings, And Power Cord; Line Voltage - Agilent Technologies 8860 Installation And First Startup

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

1
Installing the GC
Step 2 Verify line voltage, voltage settings, and
power cord
1 Locate the power label near the power cord connector on the back of the GC. Compare the
2 Verify that the power cord is correct for the voltage and location.
Electrical shock hazard. To avoid injury, only a qualified person should measure line voltage.
WAR N IN G
3 Have a qualified person measure the actual power outlet voltage and verify that it meets
The following sections detail the power specifications and requirements for reference.

Line voltage

Installation and First Startup
instrument power settings with the laboratory line voltage. See
page 16.
available"
on page 17.
the tolerance requirements listed in
"Line voltage"
on page 15.
Typical GC System - 8860 GC with computer and printer.
Leave 12-in. (31 cm)
open space for
operational access
Table 1
on page 16. See
Use isolated circuits with dedicated
grounds for GC and MSD.
31-in. (79 cm)
(without heat
deflector
26-in. (66 cm)
(with heat
deflector)
"Power consumption"
"Power cords
"Grounding"
on page
21"
on
and
15

Advertisement

Table of Contents
loading

Table of Contents