Emptying The Canister System - Weinmann ACCUVAC Pro Instructions For Use Manual

Suction device
Hide thumbs Also See for ACCUVAC Pro:
Table of Contents

Advertisement

Risk of infection due to suction material in the device!
Suction material in the device contaminates the device and
reduces its suction capacity. This can infect the patient or user.
 Always wear suitable gloves.
 Do not use the device.
 Contact a technician authorized by
10. If suction material enters the device: Contact a technician
Suction material can enter the device if the device tilts over, for
example. Reduced suction capacity is an indicator for suction
material in the device.
11. Switch off the device
12. Reprocess the device hygienically
13. Perform a function check
Result
The suction is performed.

4.9 Emptying the canister system

4.9.1 Emptying the reusable secretions canister
This section describes how the reusable secretions canister can be
emptied during suction and reused on the same patient. The
reusable canister system must be hygienically reprocessed before
being used on a new patient
the reusable canister system",
Requirement
ATMOS MedizinTechnik GmbH & Co. KG.
authorized by ATMOS MedizinTechnik GmbH & Co. KG.
page 62).
reprocessing",
page 74).
check",
page 89).
The device is switched off
page 62).
The reusable suction hose is uncoiled from the suction hose
reel and removed from the hose guide.
ACCUVAC Pro/ACCUVAC Lite
4 Preparation and operation
(see „4.7 Switching off the device",
(see „5 Hygienic
(see „6.2 Performing a function
(see „5.5 Hygienic reprocessing of
page 82).
(see „4.7 Switching off the device",
EN
65