Saving The Event Manually In The Session Data Set - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

5 Operation
Result
5.9.2 Pausing/muting the audio alarm output
Requirement
Result
5.9.3 Canceling muting or pausing of audio alarm
Requirement
Result
5.10 Saving the event manually in the
98
EN
MEDUCORE Standard
The audio alarm output is canceled for this alarm. The symbol
appears on the display and no audio signal is output for this alarm.
An alarm is active and is audible.
1. Press and hold the alarm button
The audio alarm output pauses for the time set in the operator
menu (Operator menu | Alarm settings | Pause audio). The symbol
appears on the display. If you set the time in the operator
menu to ∞ (infinite), the audio alarm output is permanently paused
(audio alarm output is muted). The symbol
display.
If set, a reminder signal will remind you at certain intervals that the
audio alarm output is paused or muted. You can set the reminder
signal in the operator menu (Operator menu | Alarm settings |
Reminder signal).
output
An alarm is active and is muted or paused.
1. Briefly (< 2 s) press the alarm button
or
Switch modes.
Muting or pausing of audio alarm output is canceled.
session data set
Measured values and the user actions performed on the device are
saved in the internal memory and on the SD card.
With the event button
the device automatically (e.g. intubation, medication
administration, etc.) can be saved in the data set in order to
subsequently assign them chronologically during evaluation.
2
(> 2 s.).
.
, events which cannot be recorded by
appears on the