11. Technical data

11.1 Specifications

Product class as per
93/42/EEC
Dimensions WxHxD in
cm
Weight
Temperature range
– operation
– storage
Adm. humidity for
operation and storage
Air pressure range
Diameter of breathing
hose connection (mask)
in mm
Electrical connection
Power consumption
during
– operation
– standby
Classification as per
prEN 60601-1:2004
– type of protection
from electric shock
– level of protection
from electric shock
– protection against
damaging ingress of
water
– operating mode
42
Technical data
CPAP 20e
21 x 9.8 x 18.5
approx. 1.25 kg
If the unit runs at +40 °C , the air output can heat up to 42 °C.
600 - 1100 hPa (automatic altitude compensation)
(allows for operation in an altitude of up to approximately 4000 meters)
19.5 (suitable for 22 mm standard cone)
115-230 V AC +10/-15 %, 50–60 Hz with WM 24480
power supply or 12-24 V DC -15/+25 % with WM 24469 DC adapter
230 V
115 V
24 V
0.1 A
0.2 A
0.8 A
0.02 A
0.03 A
0.2 A
CPAP 20e with CLICK 2
IIa
approx. 1.5 kg (without water)
+5 °C to +35 °C
–20 °C to +70 °C
≤ 95 % rF (no condensation)
12 V
230 V
1.6 A
0.14 A
0.24 A
0.02 A
Protection class II
Type B
IPX1
Constant operation
21 x 14 x 31.3
115 V
24 V
12 V
0.28 A
1.3 A
4.0 A
0.03 A
0.2 A
0.24 A

Advertisement