Technical Data; Specifications - Weinmann SOMNOvent CR Manual

Cs therapy device with autotrilevel principle
Hide thumbs Also See for SOMNOvent CR:
Table of Contents

Advertisement

12. Technical data

12.1 Specifications

Product class to
93/42/EEC
Dimensions WxHxD
in cm
Weight
Temperature range
Operation
Storage
Permitted humidity in
operation and
storage
Air pressure range
Diameter of
breathing tube
connection (mask
side) in mm
Electrical rating
Current consumption
during
operation
standby
SOMNOvent CR
18 x 9 x 32
approx. 3.4 kg
If the device is operated at +40 °C, the air given off can heat up to
≤ 95 % rh (no condensation)
600 - 1100 hPa (allows operation at altitudes of up to 4000 m)
Automatic altitude compensation
19.5 (fits 22 mm standard tapered connector)
or 12/24 V DC (with inverter WM 24131/WM 24132)
(to guarantee constant pressure, the drop in power may not
230 V 115 V
24 V
0.11 A 0.22 A 0.8 A 1.5 A
0.02 A 0.03 A 0.5 A 0.6 A
SOMNOvent CR with
IIa
approx. 3.6 kg (excl. water)
+5 °C to +35 °C
–20 °C to +70 °C
as high as 42 °C.
115/230 V AC, 50–60 Hz
exceed 10 %)
12 V
230 V 115 V
0.22 A 0.44 A 1.6 A 2.8 A
0.02 A 0.03 A 0.5 A 0.6 A
Technical data
SOMNOclick 300
18 x 9 x 44
24 V
12 V
EN