6.10 Replacing the alarm signalling device
Tools required:
•
Crosstip screwdriver, size 2,
•
Crosstip screwdriver, size 1.
1. Open the device (see "6.4 Opening the
device" on page 22).
2.
Up to appliance no.: Basic 1019; Basic p 1399
Pull the pressure measurement connection
upwards out of the housing wall.
3. Pull the connector
X5
4. Unscrew both screws
5. Remove the defective alarm signalling
device
.
23
6. Insert the new alarm signalling device
7. Secure the alarm signalling device using the
two screws
.
42
8. Push the connector
X5
circuit board.
9.
Up to appliance no.: Basic 1019; Basic p 1399
Push the pressure measurement connection
into the wall of the housing.
10. Close the device (see "6.5 Closing the
device" on page 23).
11. Perform a functional check (see "3. Final
Check" on page 8).