5.4 Alarm list
5.4.1 Storage of alarms
All alarm types listed in the tables "Physiological alarms"and "Technical alarms" are
recorded in an alarm list with date, time and duration once the alarm threshold is reached.
Up to 200 alarms can be stored. After that, the oldest alarm in each case is overwritten.
To call up the alarm list, select the menu item Alarm list in the patient menu using the dial
and confirm your selection by pressing the dial.
The alarm list is retained even if the entire power supply (power supply and internal battery)
fails.
In this case, the data can be called up for up to two years. The alarm list is overwritten after
two years or when servicing is carried out.
Alarms are always retained in the event of a power failure.
3. Use the dial to select the menu item Auto switch-on
and confirm the selection by pressing the dial. Now
select on or off using the dial. Confirm the selection
by pressing the dial. The selection bar switches back to
Auto switch-on. The current setting (on/off) is now
shown again in the Auto switch-on menu line.
4. Exit the menu again by pressing the menu key, now
assigned the back function. Auto switch-on is now
activated or deactivated.
EN
Operation
53