Controller Parameters For Communication - Thermo Scientific Lindberg/Blue M Mini-Mite TF55030 Installation And Operating Manual

Tube furnace
Table of Contents

Advertisement

| Communication Option
Chapter 7
|
7-47
304243H02

Controller Parameters for Communication

Table 4 Controller Parameters for Communication
Parameter
Comms Module Identity
Communications Protocol
Communication Interface
Baud Rate
Parity
Comms Address
Troubleshooting Communications
If your connection is not working properly, check the following conditions:
A. Verify complete and tight cable connections between the furnace and the PC.
B. Verify that power has been supplied to the unit and temperature controller before
starting the software program.
c. Verify DB-9 pin connections as shown in wiring diagram, interchange 2 & 3
connections if communication is not working.
d. Verify the configuration values in the controller, listed in the Table 4, "Controller
Parameters for Communication".
Value
Comms (67)
Modbus
RS 485
9600_baud (0)
none
1
Mini-mite Tube Furnace

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lindberg/blue m mini-mite tf55035

Table of Contents