Left Hand Front Panel And Left Hand Sub Panel - Thermo Scientific TSQ 8000 Evo Hardware Manual

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Covers of the TSQ 8000 Evo Instrument

Left Hand Front Panel and Left Hand Sub Panel

256
TSQ 8000 Evo Hardware Manual
To replace the left hand front panel and the left hand sub panel of the TSQ 8000 Evo
instrument
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107. You do not have to remove the ion source cartridge or column or
vent the system. If you are using hydrogen as a carrier gas, power off the GC.
2. Remove the four screws on the left hand front panel. One is located on the bottom left of
the interior front panel of the instrument.
Figure 273. Locating the Left Hand Front Panel Screws
3. To replace the left hand sub panel, disconnect the inputs as shown in
"Shutting Down the TSQ 8000 Evo
Screws
Screw on Front
of Instrument
Figure
274.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents