Advanced Troubleshooting
Replacing Components of the Ion Source

Source Heater Block, Lens Heater Block, and Lens/Source Heater

142
TSQ 8000 Evo Hardware Manual
To replace the source heater block, lens heater block, or lens/source heater
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering these
components.
WARNING - INSTRUMENT DAMAGE Thermo Fisher Scientific strongly recommends
that only trained Thermo Fisher Scientific Field Service Engineers remove and service
these components.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. Disconnect the Q1 entrance lens, Q2 entrance lens, Q3 entrance lens, and Q3 exit lens
wires from the source interface board using a flat-tipped screwdriver.
Figure 118. Locating Wires on the Source Interface Board
"Shutting Down the TSQ 8000 Evo
on
page
109.