Step 3:
To remove the front panel from the chassis
Removing the
1. Disconnect the solvent lines from the proportioning valves.
Front Panel
2. Using an 8-mm open-end wrench, disconnect the crossover tube from the purge valve and the left side of
the LDA.
3. Pull the tubing off the purge valve and the pulse dampener flush valve.
4. Being careful not to remove the spigots, remove the wingnuts from the purge valve and the pulse dampener
flush valve.
Spigots
5. Using a needle nose pliers, squeeze the back of the clip that secures the solvent tubing and push the clip forward
to remove it from the front panel. As shown below, the clip is located in front of the solvent degassing unit.
Set the clip aside. When you reassemble the pump, install the clip in the new front panel.