Appendix O: Advanced Applications
KCON setup
After connecting the system as indicated in
(KCON) to add the switch matrix and test fixture to the system configuration. KCON is used to
manage the configuration of all instrumentation controlled by the 4200-SCS software. Once the
matrix and test fixture have been added, and the instrument-to-matrix-to-pins connections have
been defined, specify an instrument terminal and test system pin and KITE will automatically
connect the two using the matrix. Changes to the system configuration will only be necessary
when changes to instrument-to-matrix-to-pins wiring are made.
Follow the steps below to start KCON and modify the system configuration as described above.
For additional information regarding KCON, refer to the
Section
1.
On the desktop, double-click the KCON icon to open KCON.
2.
Using the Tools menu, add a switch matrix to the system configuration as indicated in
Figure
Figure O-2
Add a switch matrix to the system configuration
3.
Using the Tools menu, add a test fixture to the system configuration as indicated in
Figure
Figure O-3
Add a test fixture to the system configuration
O-4
7.
O-2.
O-3.
Return to
Figure
O-1, run the Keithley CONfiguration utility
Keithley CONfiguration Utility (KCON),
Section Topics
Model 4200-SCS Reference Manual
4200-901-01 Rev. S / May 2017