Model 4200-SCS Reference Manual
NOTE:
Figure L-4
Select Host Interface
Figure L-5
Signatone GPIB driver window
NOTE:
Figure L-6
Set GPIB Address
4200-901-01 Rev. S / May 2017
b. Select the IEEE488 Host Interface.
If IEEE488 has been enabled, press Cancel and skip the GPIB setup.
c. The Signatone GPIB driver window will show on the screen.
d. Press Addr and check that the GPIB address matches the GPIB_Address setting in the
S4200-SCS prober configuration file, prbcnfg_CM500.dat located at
C:\S4200\sys\dat.
The default GPIB address is set to 28.
e. If the address does not match, enter the new GPIB address, then press OK.
Return to
Section Topics
Appendix L: Signatone CM500 Prober
L-5