Model 4200-SCS Reference Manual
Figure K-1
Sample SUMMIT-12K prober configuration file
# prbcnfg_CC12K.dat - DEFAULT Prober Configuration
File
#
# The following tag, "PRBCNFG", is used by the
engine in order to determine
# the MAX number of SLOTS and CASSETTES for a given
prober at runtime.
#
<PRBCNFG>
#
# for OPTIONS "" == NULL, max 32 chars in string
#
# Example
# 01234567890
#PROBER_1_OPTIONS=1,1,1,1,1,1
#
#
#OcrPresent
#AutoAlnPresent
#ProfilerPresent
#HotchuckPresent
#HandlerPresent
#Probe2PadPresent
#
#
# Configuration for direct GPIB probers:
# CC12K
#
PROBER_1_PROBTYPE=CC12K
PROBER_1_OPTIONS=0,0,0,0,1,0
PROBER_1_IO_MODE=GPIB
PROBER_1_GPIB_UNIT=0
PROBER_1_GPIB_SLOT=1
PROBER_1_GPIB_ADDRESS=28
PROBER_1_GPIB_WRITEMODE=0
PROBER_1_GPIB_READMODE=2
PROBER_1_GPIB_TERMINATOR=13
PROBER_1_TIMEOUT=300
PROBER_1_SHORT_TIMEOUT=5
Step 1. Set up communication
The Cascade SUMMIT-12000 prober is configured for GPIB communication only. Ensure that the
prober configuration file is set up properly for the GPIB communication interface (see
the prober configuration file earlier in this
NOTE:
1.
4200-901-01 Rev. S / May 2017
The following configuration is accomplished using the probe station PC.
Connect the Model 4200-SCS GPIB port and the probe station PC's GPIB port using a
shielded IEEE-488 cable (Model 7007). Refer to
to
Figure K-3
for a connection diagram.