Appendix N: Additional User Libraries
RETURNED STATUS VALUES:
Returned values are placed in the spread sheet (Sheet tab).
0
-10030
-10090
-10100
-10102
OpenCal4294 user module
User module description
The OpenCal4294 routine performs OPEN calibration for the HP 4294 IMP meter.
shows the default parameters for the OpenCal4294 user module. After OPEN calibration is
performed, it should then be verified by performing an FiSweep on the calibration device.
Syntax:
status = OpenCal4294(char *Inststr, int Adaptor, double OpenRefG, double OpenRefC);
Figure N-4
OpenCal4294 user module (default parameters)
INPUTS:
InstldStr
Adaptor
OpenRefG
OpenRefC
N-8
OK.
HP 4294 not in KCON.
GPIB time-out occurred during communications.
An invalid input parameter is specified.
There is an error when parsing the HP 4294's response.
(char *) KCON instrument ID. Default is CMTR1.
(int) Type of cable adapter. Valid input is 1 to 4:
1
Specifies Agilent 16048G (2 pairs of 1-meter coaxial cables).
2
Specifies Agilent 16048H (2 pairs of 2-meter coaxial cables).
3
Specifies Agilent 42942A (7-millimeter 42942A probe).
4
Specifies Agilent 42941A (Advanced impedance probe).
(double) Reference conductance value of the OPEN.
(double) Reference capacitance value of the OPEN.
Return to
Section Topics
Model 4200-SCS Reference Manual
Figure N-4
4200-901-01 Rev. S / May 2017