Keithley 4200-SCS Reference Manual page 1673

Semiconductor characterization system
Hide thumbs Also See for 4200-SCS:
Table of Contents

Advertisement

Model 4200-SCS Reference Manual
Syntax:
status = HotChuck Temp(int GPIBAddress, double TargetTemp);
Figure N-8
SetChuckTemp user module (default parameters)
INPUTS:
TargetTemp
RETURNED STATUS VALUES:
Returned values are placed in the spread sheet (Sheet tab).
0
-2
-10000
-10090
-10091
Triotek receive commands:
All receive commands have character R as the leading character. "R" commands are used with a
user provided computer with IEEE for remote control of the TC controller. Below is a list of receive
commands and their descriptions:
RAn Command
RBn Command
RDxx.x Command
RSxxx.x Command Used to set temperature.
RPn Command
RNn Command
4200-901-01 Rev. S / May 2017
(double) Target temperature (in ºC) for the chuck.
OK.
Unable to set temperature.
(INVAL_INST_ID) The specified instrument ID does not exist.
(GPIB_ERROR_OCCURRED) A GPIB communications error occurred.
(GPIB_TIMEOUT) A time-out occurred during communications.
Used to set beginning ramp number.
n = 1 to 8
Used to activate a preset temperature.
n = 1 to 8
Used to set Dewpoint Limit.
-20.0 ≤ xx.x ≤ 20.0
Lower Temp Limit ≤ xxx.x ≤ Upper Temp Limit
Used to inhibit chuck power.
n = 0 power inhibited
n = 1 power enabled
Used to set program cycle.
1 ≤ n ≤ 999
Return to
Section Topics
Appendix N: Additional User Libraries
N-13