Model 4200-SCS Reference Manual
Figure I-47
KITE: probesites project
2.
3.
probesubsites KITE project example
The following is a step by step procedure to properly configure the 8860 so the probesubsites
KITE project executes successfully.
NOTE:
When using pcIndie, ensure that the project and the program listing on the Micromanipulator
match (the program listing is a list of absolute chuck moves in the order of execution). When
creating the program listing, use a repeatable pattern.
Example
To configure the 8860:
4200-901-01 Rev. S / May 2017
Select the top node in the ProjectView tab of the Project Navigator window.
Click the green Run button.
The following configuration is accomplished using the pcBridge computer.
Five dies have been selected for probing
been selected.
1) Move to the first subsite of the first die
2) Add it to the program listing
3) Move to the second subsite on the first die
4) Add it to the program listing
5) Move to the first subsite on the second die
6) Add it to the program
7) Continue moving and adding until all subsites have been entered into the list
Return to
(Figure
Section Topics
Appendix I: Micromanipulator 8860 Prober
I-48). On each die, two sub-sites have