Appendix N: Additional User Libraries
Figure N-2
FiSweep4294 user module (default parameters)
INPUTS:
InstldStr
Adaptor
StartF
StopF
Number_of_point (int) Number of points in the voltage sweep. Valid range of inputs is 8 to 801
Osc_type
Power_Level
Model
BandWidth
Csize, Vsize,
Thetasize
N-6
(char *) KCON instrument ID. Default is CMTR1.
(int) Type of cable adapter. Valid input is 1 to 4:
1Specifies Agilent 16048G (2 pairs of 1-meter coaxial cables).
2Specifies Agilent 16048H (2 pairs of 2-meter coaxial cables).
3Specifies Agilent 42942A (7-millimeter 42942A probe).
4Specifies Agilent 42941A (Advanced impedance probe).
(double) Starting frequency of the sweep. Valid range of inputs is 40 to 1.1E+8
Hz volts.
(double) Ending frequency of the sweep. Valid range of inputs is 40 to 1.1E+8
Hz volts.
points.
(int) Type of AC stimulus: 0 = voltage, 1 = current. Default is voltage.
(double) OSC source level. Valid inputs are 0.005 to 1 volt.
(int) Measurement model. Supported models include the following:
CsRs
CpG
R-X
G-B
Z-Theta
CpD
(int) Measurement bandwidth factor. Valid inputs are from 1 to 5.
(int) Must be equal to or greater than the specified Number_of_point.