Appendix E: Status model
Status function summary
The following functions and attributes control and read the various registers. Additional information for
the various register sets is included later in this section. Also, refer to the specific command as listed
in
TSP commands
Status function summary
Type
Status summary
Measurement event
Operation status
Questionable status
Standard event
System summary
* = .condition, .event, .ntr, .ptr and .enable
smuX: For Models 2601B, 2611B, and 2635B, this value is smua (SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be smua (for SMU Channel A) or smub (for SMU Channel B).
E-4
Series 2600B System SourceMeter® Instrument Reference Manual
(on page 7-7).
Function or attribute
status.condition
status.node_enable
status.node_event
status.request_enable
status.request_event
status.reset
status.measurement.*
status.measurement.buffer_available.*
status.measurement.current_limit.*
status.measurement.instrument.*
status.measurement.instrument.smuX.*
status.measurement.reading_overflow.*
status.measurement.voltage_limit.*
status.operation.*
status.operation.calibrating.*
status.operation.instrument.*
status.operation.instrument.digio.*
status.operation.instrument.digio.trigger_overrun.*
status.operation.instrument.lan.*
status.operation.instrument.lan.trigger_overrun.*
status.operation.instrument.smuX.*
status.operation.instrument.smuX.trigger_overrun.*
status.operation.instrument.trigger_blender.*
status.operation.instrument.trigger_blender.trigger_overrun.*
status.operation.instrument.trigger_timer.*
status.operation.instrument.trigger_timer.trigger_overrun.*
status.operation.instrument.tsplink.*
status.operation.instrument.tsplink.trigger_overrun.*
status.operation.measuring.*
status.operation.remote.*
status.operation.sweeping.*
status.operation.trigger_overrun.*
status.operation.user.*
status.questionable.*
status.questionable.calibration.*
status.questionable.instrument.*
status.questionable.instrument.smuX.*
status.questionable.over_temperature.*
status.questionable.unstable_output.*
status.standard.*
status.system.*
status.system2.*
status.system3.*
status.system4.*
status.system5.*
2600BS-901-01 Rev. B / May 2013