Section 7: TSP command reference

setup.recall()

This function recalls settings from a saved setup.
Type
TSP-Link accessible
Function
Yes
Usage
setup.recall(id)
id
Details
When the id parameter is an integer (n), it is interpreted as the setup number to restore from the instrument's
nonvolatile memory. When n = 0, the instrument recalls the factory default setup; when n = 1 to 5, the
instrument recalls a user-saved setup.
When the id parameter is a string, it is interpreted as the path and file name of the setup to restore from a file on
a USB flash drive. The path may be absolute or relative to the current working directory.
Before a setup is recalled, an instrument reset is performed.
Example 1
setup.recall(1)
Example 2
setup.recall("/usb1/KEITHLEY_30730.set")
Also see
setup.save()
(on page 7-181)
Saved setups
7-180
Affected by
An integer or string that specifies the location of the setup to recall:
Factory default setup (0)
User-saved setup in nonvolatile memory (1 to 5)
User-saved setup on a USB flash drive ("/path/filename")
(on page 2-46)
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Recall the user-saved setup at location 1.
Recall a user-saved setup stored in a file
named KEITHLEY_30730 on a USB flash
drive.
Default value
2600BS-901-01 Rev. B / May 2013