Smux.contact.check() - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example
-- Short SENSE LO and LO terminals
-- Short SENSE HI and HI terminals
-- Allow readings to settle, then get measurements
r0_hi, r0_lo = smua.contact.r()
-- Connect 50 OHM resistor between SENSE LO and LO
-- Connect 50 OHM resistor between SENSE HI and HI
-- Allow readings to settle, then get measurements
r50_hi, r50_lo = smua.contact.r()
smua.contact.calibratelo(r0_lo, Z_actual_lo, r50_lo,
50_ohm_actual_lo)
smua.contact.calibratehi(r0_hi, Z_actual_hi, r50_hi,
50_ohm_actual_hi)
Also see
Adjustment
smuX.cal.restore()
smuX.cal.save()
smuX.cal.unlock()
smuX.contact.calibratehi()

smuX.contact.check()

This function determines if contact resistance is lower than the threshold. This function is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.contact.check()
X
2600BS-901-01 Rev. B / May 2013
(on page B-18)
(on page 7-188)
(on page 7-189)
(on page 7-190)
(on page 7-191)
Affected by
Source-measure unit (SMU) channel (for example, smua.contact.check()
applies to SMU channel A)
Section 7: TSP command reference
Where saved
Default value
Performs contact
check on SMU
channel A.
Install and measure
two resisters.
The user sends
contact check LO
calibration
command.
The user sends
contact check HI
calibration
command.
7-193

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents