Section 7: TSP command reference
smuX.measure.rel.levelY
This attribute sets the offset value for relative measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
relValue = smuX.measure.rel.levelY
smuX.measure.rel.levelY = relValue
relValue
X
Y
Details
This attribute specifies the offset value used for relative measurements. When relative measurements are
enabled (see smuX.measure.rel.enableY), all subsequent measured readings will be offset by the value of
this attribute. Each returned measured relative reading is the result of the following calculation:
Relative reading = Actual measured reading - Relative offset value
Example
smua.measure.rel.levelv = smua.measure.v()
Also see
Relative offset
reset()
(on page 7-160)
smuX.measure.rel.enableY
smuX.reset()
(on page 7-216)
7-212
Affected by
Instrument reset
SMU reset
Recall setup
Relative measurement offset value
Source-measure unit (SMU) channel (for example, smua.measure.rel.levelv
applies to SMU channel A)
SMU measurement function (v = voltage, i = current, r = resistance, p = power)
(on page 3-1)
(on page 7-211)
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Not saved
Performs a voltage measurement using
SMU channel A and then uses it as the
relative offset value.
Default value
0
2600BS-901-01 Rev. B / May 2013