Lan.trigger[N].Wait(); Localnode.autolinefreq - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example
lan.trigger[5].stimulus = trigger.timer[1].EVENT_ID
Also see
lan.trigger[N].assert()
lan.trigger[N].clear()
lan.trigger[N].connect()
lan.trigger[N].overrun

lan.trigger[N].wait()

lan.trigger[N].wait()
This function waits for an input trigger.
Type
TSP-Link accessible
Function
Yes
Usage
triggered = lan.trigger[N].wait(timeout)
triggered
N
timeout
Details
If one or more trigger events have been detected since the last time lan.trigger[N].wait() or
lan.trigger[N].clear() was called, this function returns immediately.
After waiting for a LAN trigger event with this function, the event detector is automatically reset and rearmed
regardless of the number of events detected.
Example
triggered = lan.trigger[5].wait(3)
Also see
lan.trigger[N].assert()
lan.trigger[N].clear()
lan.trigger[N].overrun
lan.trigger[N].stimulus

localnode.autolinefreq

This attribute enables or disables automatic power line frequency detection at start-up.
Type
TSP-Link accessible
Attribute (RW)
Yes
2600BS-901-01 Rev. B / May 2013
(on page 7-128)
(on page 7-129)
(on page 7-130)
(on page 7-134)
(on page 7-137)
Affected by
Trigger detection indication
The trigger packet over LAN to wait for (1 to 8)
Maximum amount of time in seconds to wait for the trigger event
(on page 7-128)
(on page 7-129)
(on page 7-134)
(on page 7-135)
Affected by
Not applicable
Section 7: TSP command reference
Use timer 1 trigger event as
the source for LAN packet 5
trigger stimulus.
Where saved
Default value
Wait for a trigger with LAN packet 5 with a timeout of
3 seconds.
Where saved
Default value
Nonvolatile memory
true (enabled)
7-137