Tsplink.trigger[N].Pulsewidth; Tsplink.trigger[N].Release() - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference
Example
print(tsplink.trigger[1].overrun)
Also see
tsplink.trigger[N].assert()
tsplink.trigger[N].clear()
tsplink.trigger[N].mode

tsplink.trigger[N].release()

tsplink.trigger[N].reset()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()

tsplink.trigger[N].pulsewidth

This attribute sets the length of time that the trigger line is asserted for output triggers. This attribute is not
available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
width = tsplink.trigger[N].pulsewidth
tsplink.trigger[N].pulsewidth = width
width
N
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6 Sets pulse width for trigger line 3 to 20 μs.
Also see
tsplink.trigger[N].release()
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line. This function is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Function
Yes
Usage
tsplink.trigger[N].release()
N
The trigger line (1 to 3)
7-360
Series 2600B System SourceMeter® Instrument Reference Manual
(on page 7-356)
(on page 7-356)
(on page 7-358)
(on page 7-360)
(on page 7-361)
(on page 7-362)
(on page 7-364)
Affected by
Instrument reset
TSP-Link trigger N reset
Recall setup
The pulse width (in seconds)
The trigger line (1 to 3)
(on page 7-360)
Affected by
If an event was ignored, displays
true; if an event was not ignored,
displays false.
Where saved
Default value
10e-6 (10 µs)
Not saved
Where saved
Default value
2600BS-901-01 Rev. B / May 2013