Section 3: Functions and features
Series 2600B System SourceMeter® Instrument Reference Manual
2. To run the i_leakage_measure() function in the KIHighC factory script, send:
-- Charges the capacitor.
smua.source.levelv = 5
smua.source.output = smua.OUTPUT_ON
delay(1)
imeas = i_leakage_measure(smua, 0, 1, 300e-3, 10e-6, 100e-3)
-- The parameters in the i_leakage_measure() function represent
-- the following:
-- smu = smua
-- levelv = 0 V
-- limiti = 1 A
-- sourcedelay = 300 ms
-- measurei = 10 uA range
-- measuredelay = 100 ms
Adjust the voltage level and source delays based on the value and type of capacitor along with the
magnitude of the voltage step and the current measure range.
Figure 85: Enabling high-capacitance mode
3-70
2600BS-901-01 Rev. B / May 2013