Series 2600B System SourceMeter® Instrument Reference Manual
Also see
Measure auto delay
reset()
(on page 7-160)
smuX.measure.count
smuX.measure.delayfactor
smuX.source.delay
smuX.reset()
smuX.measure.delayfactor
This attribute stores a multiplier to the delays that are used when smuX.measure.delay is set to
smuX.DELAY_AUTO.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
delayFactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayFactor
delayFactor
X
Details
The delay factor is only applied when smuX.measure.delay = smuX.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic delay.
Setting this attribute to zero disables delays when smuX.measure.delay = smuX.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Also see
Measure auto delay
reset()
(on page 7-160)
smuX.measure.delay
smuX.reset()
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type
TSP-Link accessible
Attribute (RW)
Yes
2600BS-901-01 Rev. B / May 2013
(on page 2-83)
(on page 7-201)
(on page 7-203)
(see "<sm.source.delay" on page 7-220)
(on page 7-216)
Affected by
Instrument reset
SMU reset
Recall setup
The delay factor multiplier
Source-measure unit (SMU) channel (for example, smua.measure.delayfactor
applies to SMU channel A)
(on page 2-83)
(on page 7-202)
(on page 7-216)
Affected by
Instrument reset
SMU reset
Recall setup
Section 7: TSP command reference
Where saved
Default value
Not saved
1
Doubles the measure delay for SMU channel A.
Where saved
Default value
Saved setup
1
7-203