trigger.timer[N].delay

This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
interval
N
Details
Each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[N].delaylist = {interval}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
Example
trigger.timer[1].delay = 50e-6
Also see
trigger.timer[N].reset()

trigger.timer[N].delaylist

This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
intervals = trigger.timer[N].delaylist
trigger.timer[N].delaylist = intervals
intervals
N
Details
Each time the timer is triggered, it uses the next delay period from the array. The default value is an array with
one value of 10 µs.
After all elements in the array have been used, the delays restart at the beginning of the list.
If the array contains more than one element, the average of the delay intervals in the list must be ≥ 50 µs.
2600BS-901-01 Rev. B / May 2013
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Delay interval in seconds (1.00e-09 to 100,000)
Trigger timer number (1 to 8)
(on page 7-348)
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Table of delay intervals in seconds
Trigger timer number (1 to 8)
Section 7: TSP command reference
Where saved
Default value
10e-6 (10 µs)
Not saved
Set the trigger timer 1 to delay for
50 µs.
Where saved
Default value
Not saved
{10e-6}
7-345

Advertisement

Table of Contents
loading

This manual is also suitable for:

2604b2614b2634b2635b2636b2602b ... Show all

Table of Contents