Section 7: TSP command reference
Usage
func = display.smuX.limit.func
display.smuX.limit.func = func
func
X
Details
Selects the displayed limit function: primary (IV) or power (P).
SMU A and SMU B can be set for different display functions.
Example
display.smua.limit.func = display.LIMIT_P
Also see
display.smuX.measure.func
display.smuX.measure.func
This attribute specifies the type of measurement that is being displayed.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
func = display.smuX.measure.func
display.smuX.measure.func = func
func
X
Details
Selects the displayed measurement function: Amperes, volts, ohms, or watts.
SMU A and SMU B can be set for different measurement functions.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
7-78
Series 2600B System SourceMeter® Instrument Reference Manual
One of the following values:
0 or display.LIMIT_IV: Displays the primary limit setting
1 or display.LIMIT_P: Displays the power limit setting
Source-measure unit (SMU) channel (for example, display.smua.limit.func
applies to SMU channel A)
(on page 7-78)
Affected by
Instrument reset
Recall setup
One of the following values:
•
0 or display.MEASURE_DCAMPS: Selects current measurement function
•
1 or display.MEASURE_DCVOLTS: Selects volts measurement function
•
2 or display.MEASURE_OHMS: Selects ohms measurement function
•
3 or display.MEASURE_WATTS: Selects power measurement function
Source-measure unit (SMU) channel (for example,
display.smua.measure.func applies to SMU channel A)
Specifies the power limit value is
displayed for SMU Channel A.
Where saved
Default value
Saved setup
1 (display.MEASURE_DCVOLTS)
Selects the current measure function
for SMU A.
2600BS-901-01 Rev. B / May 2013