Smux.measure.filter.enable - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Section 7: TSP command reference
Usage
filterCount = smuX.measure.filter.count
smuX.measure.filter.count = filterCount
filterCount
X
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Also see
Filters
(on page 3-3)
reset()
(on page 7-160)
setup.recall()

smuX.measure.filter.enable

smuX.measure.filter.type
smuX.reset()
(on page 7-216)
smuX.measure.filter.enable
This command enables or disables filtered measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
filterState = smuX.measure.filter.enable
smuX.measure.filter.enable = filterState
filterState
X
Details
This command enables or disables the filter.
Example
smua.measure.filter.enable = 1
7-204
The number of readings required for each filtered measurement (1 to 100)
Source-measure unit (SMU) channel (for example,
smua.measure.filter.count applies to SMU channel A)
(on page 7-180)
(on page 7-204)
(on page 7-205)
Affected by
Instrument reset
SMU reset
Recall setup
The filter status; set to one of the following values:
0 or smuX.FILTER_OFF: Disables the filter
1 or smuX.FILTER_ON: Enables the filter
SMU channel (for example, smua.measure.filter.enable applies to SMU
channel A)
Series 2600B System SourceMeter® Instrument Reference Manual
Sets the filter count for SMU channel A to 10.
Where saved
Saved setup
Enables the filter for SMU channel A. Alternatively, the
value 1 may be replaced with smua.FILTER_ON.
Default value
0 (smuX.FILTER_OFF)