Series 2600B System SourceMeter® Instrument Reference Manual
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.blender[N].reset()
N
Details
The trigger.blender[N].reset() function resets the following attributes to their factory defaults:
•
trigger.blender[N].orenable
•
trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Example
trigger.blender[1].reset()
Also see
trigger.blender[N].orenable
trigger.blender[N].overrun
trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = eventID
eventID
N
M
Details
There are four stimulus inputs that can each select a different event. The eventID parameter can be the event
ID of any trigger event.
Use zero to disable the blender input.
2600BS-901-01 Rev. B / May 2013
Affected by
The trigger event blender (1 to 6)
(on page 7-337)
(on page 7-338)
(on page 7-339)
Affected by
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
The event that triggers the blender action; see Details
An integer representing the trigger event blender (1 to 6)
An integer representing the stimulus index (1 to 4)
Section 7: TSP command reference
Where saved
Default value
Resets the trigger blender 1
settings to factory defaults.
Where saved
Default value
Configuration script trigger.EVENT_NONE
7-339