Section 7: TSP command reference
Details
This function resets the following attributes to factory default settings:
•
digio.trigger[N].mode
•
digio.trigger[N].pulsewidth
•
digio.trigger[N].stimulus
It also clears digio.trigger[N].overrun.
Example
digio.trigger[3].mode = 2
digio.trigger[3].pulsewidth = 50e-6
digio.trigger[3].stimulus = digio.trigger[5].EVENT_ID
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
digio.trigger[3].reset()
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
Set the digital I/O trigger line 3 for a falling edge with a pulsewidth of 50 microseconds.
Use digital I/O line 5 to trigger the event on line 3.
Reset the line back to factory default values.
Output before reset:
2.00000e+00
Output after reset:
0.00000e+00
Also see
digio.trigger[N].mode
digio.trigger[N].overrun
digio.trigger[N].pulsewidth
digio.trigger[N].stimulus
digio.trigger[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line. This attribute is not
available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
triggerStimulus = digio.trigger[N].stimulus
digio.trigger[N].stimulus = triggerStimulus
triggerStimulus
N
Details
Set this attribute to zero (0) to disable the automatic trigger output.
Do not use the stimulus attribute for generating output triggers under script control. Use
digio.trigger[N].assert() instead.
7-56
Series 2600B System SourceMeter® Instrument Reference Manual
5.00000e-05
1.00000e-05
(on page 7-52)
(on page 7-54)
(on page 7-54)
(on page 7-56)
Affected by
Instrument reset
Digital I/O trigger N reset
Recall setup
The event identifier for the triggering event
Digital I/O trigger line (1 to 14)
5.00000e+00
0.00000e+00
Where saved
Default value
Not saved
0