Smux.trigger.armed_Event_Id; Smux.trigger.autoclear - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual

smuX.trigger.ARMED_EVENT_ID

This constant contains the armed event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = smuX.trigger.ARMED_EVENT_ID
eventID
X
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to armed
events from this SMU.
Example
trigger.timer[1].stimulus =
smua.trigger.ARMED_EVENT_ID
Also see
Triggering

smuX.trigger.autoclear

This attribute turns automatic clearing of the event detectors on or off.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
autoClear = smuX.trigger.autoclear
smuX.trigger.autoclear = autoClear
autoClear
X
Details
This attribute enables or disables automatic clearing of the trigger model state machine event detectors when the
SMU transitions from the arm layer to the trigger layer.
Only the detected states of the event detectors are cleared.
The overrun statuses of the event detectors are not automatically cleared when the SMU transitions from the arm
layer to the trigger layer.
The event detectors are always cleared when a sweep is initiated.
2600BS-901-01 Rev. B / May 2013
Affected by
The armed event number
Source-measure unit (SMU) channel (for example,
smua.trigger.ARMED_EVENT_ID applies to SMU channel A)
(on page 3-32)
Affected by
SMU reset
Instrument reset
Recall setup
Auto clear setting; set to one of the following values:
0 or smuX.DISABLE: Turns off automatic clearing of the event detectors
1 or smuX.ENABLE: Turns on automatic clearing of the event detectors
Source-measure unit (SMU) channel (for example, smua.trigger.autoclear
applies to SMU channel A)
Section 7: TSP command reference
Where saved
Default value
Trigger timer when the SMU
passes through the ARM layer.
Where saved
Default value
Not saved
0 (smuX.DISABLE)
7-235