localnode.showerrors

This attribute sets whether or not the instrument automatically sends generated errors.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
errorMode = localnode.showerrors
localnode.showerrors = errorMode
errorMode
Details
If this attribute is set to 1, the instrument automatically sends any generated errors stored in the error queue, and
then clears the queue. Errors are processed after executing a command message (just before issuing a prompt,
if prompts are enabled).
If this attribute is set to 0, errors are left in the error queue and must be explicitly read or cleared.
When using this command from a remote node, localnode should be replaced with the node reference, for
example, node[5].showerrors.
Example
localnode.showerrors = 1
Also see
localnode.prompts

makegetter()

This function creates a function to get the value of an attribute.
Type
TSP-Link accessible
Function
No
Usage
getter = makegetter(table, attributeName)
getter
table
attributeName
2600BS-901-01 Rev. B / May 2013
Affected by
Power cycle
Enables (1) or disables (0) the show errors state
(on page 7-141)
Affected by
The return value
Read-only table where the attribute is located
A string representing the name of the attribute
Section 7: TSP command reference
Where saved
Default value
Not saved
0 (disabled)
Enables sending of generated errors.
Where saved
Default value
7-145