Localnode.reset(); Localnode.revision - Keithley 2601B Reference Manual

2600b series system sourcemeter instrument
Hide thumbs Also See for 2601B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Details
When set to 1, the IEEE Std 488.2 common commands generate prompts if prompting is enabled with the
localnode.prompts attribute. If set to 1, limit the number of *trg commands sent to a running script to 50
regardless of the setting of the localnode.prompts attribute.
When set to 0, IEEE Std 488.2 common commands will not generate prompts. When using the *trg command
with a script that executes trigger.wait() repeatedly, set localnode.prompts4882 to 0 to avoid
problems associated with the command interface input queue filling.
This attribute resets to the default value each time the instrument power is cycled.
When using this command from a remote node, localnode should be replaced with the node reference, for
example node[5].prompts4882.
Example
localnode.prompts4882 = 0
Also see
localnode.prompts

localnode.reset()

This function resets the local node instrument.
Type
TSP-Link accessible
Function
Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the node[X].reset() command.
A local node reset includes:
Source-measure unit (SMU) attributes affected by a SMU reset are reset
Other settings are restored back to factory default settings
A localnode.reset() is different than a reset() because reset() resets the entire system.
When using this command from a remote node, localnode should be replaced with the node reference, for
example node[5].reset().
Example
localnode.reset()
Also see
reset()
(on page 7-160)
smuX.reset()

localnode.revision

This attribute stores the firmware revision level.
Type
TSP-Link accessible
Attribute (R)
Yes
2600BS-901-01 Rev. B / May 2013
(on page 7-141)
Affected by
(on page 7-216)
Affected by
Not applicable
Section 7: TSP command reference
Disables IEEE Std 488.2 common command prompting.
Where saved
Default value
Resets the local node.
Where saved
Default value
Not applicable